Partial discharge
The laboratory performs the calibration in accordance with IEC 60270 by comparison measurements against standards.
The output of the calibration object is calibrated complying with IEC 60270:2000/AMD1:2015 (High-voltage test techniques - Partial discharge measurements) clauses A.2 and A.3.
The measurement uncertainty is estimated in accordance with EA-4/02 M: 2022 („Evaluation of the Uncertainty of Measurement in Calibration“).
It is checked whether the measurement uncertainty lies within the limits in accordance with IEC 60270.